Accu-Chek NANO Informacje Techniczne Strona 20

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 19
18
About the Control Solution
Only use the ACCU‑CHEK SmartView control solution.
Close the control solution bottle tightly after use.
Write the date you open the control solution bottle on the bottle label. The control solution must be
discarded 3 months from the date the bottle was opened (discard date) or on the Use By date on the
bottle label, whichever comes first.
Do not use control solution that is past the Use By or discard date.
Refer to the control solution package insert for control solution storage conditions.
The meter automatically recognizes the dierence between the control solution and blood.
The control results are not displayed in memory.
The control solution can stain fabric. Remove stains by washing with soap and water.
Przeglądanie stron 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 203 204

Komentarze do niniejszej Instrukcji

Brak uwag