Symbols, error messages and troubleshooting
195
Symbols, error messages and troubleshooting
12
Error message Meaning
" There is no test strip drum in the
meter.
Press the Test button to turn the meter
off. Insert a new test strip drum (see
Chapter 2.3).
" The bar code on the test strip drum is
damaged.
You can no longer perform tests with
test strips from this drum. Insert a
new drum.
" You inserted a test strip drum from a
different blood glucose measuring
system.
Press the Test button to turn the meter
off. Insert an Accu-Chek Compact test
strip drum (see Chapter 2.3).
Error message Meaning
In memory: the result in memory cannot
be displayed.
All other results in memory can continue
to be retrieved with the M or S button.
A meter error has occurred.
Turn the meter off and then on again by
pressing the Test button. If the error mes-
sage persists, please contact the
Accu-Chek Customer Careline free on
0800 701000 (UK) or 1800 709600
(Ireland).