Symbols, error messages and troubleshooting
181
Symbols, error messages and troubleshooting
12
Symbol Meaning

the date that it is set is in the past,
you changed the time and date,
then performed a test and after-
wards re-set the time and date, but
have not performed any more tests
since then.
The result is lower than 0.6 mmol/L
(10 mg/dL).
The result is greater than 33.3 mmol/L
(600 mg/dL).
You have turned the meter on in order to
download the results.
The meter is downloading results.
Decimal point displayed by mmol/L
meters and is part of the result (e.g. 8.2
mmol/L is displayed as

)
12.2 Interferences and error messages – causes and trouble-
shooting recommendations
If you receive frequent error messages, please contact the Accu-Chek
Customer Careline free on 0800 701000 (UK) or 1800 709600 (Ireland).
If this happens … it means:
The meter will not turn
on.
" The batteries (normal or rechargeable)
are empty or none are inserted.
Insert batteries (new or recharged).
" You inserted the batteries the wrong
way round.
Remove the batteries and re-insert
them as shown in the battery com-
partment. If the batteries were in the
meter the wrong way round for longer
than two minutes, the year, time and
date are now lost. Re-enter them (see
Chapter 3).
