R. For custom comment, press icon to enter other or additional pertinent
customized comment(s), not listed in the pre-programmed comments.
S. Once you have selected and/or entered the desired comment(s), Press [] to
return to the Results Screen.
T. Press [] again to confirm the result and send the result from the meter
wirelessly.
U. Remove the strip and discard in a biohazard waste container.
V. Repeat steps E through U to run the second level of control.
W. Proceed with patient testing, or press the Power OFF button to turn the meter off.
X. Placing the meter in the docking station ensures downloading of QC information
to the POCT Lab (back-up for the wireless data transfer) and charges the battery
at the same time.
X. PATIENT TESTING PROCEDURE
A. Patient Preparation
1. For Fingerstick and Routine Venipuncture Sample Procedure, see online
Specimen Collection Manual at:
http://labmed.ucsf.edu/labmanual/mftlngmtzn/dnld/speccolPhleboto
my_Venipuncture.pdf
2. For capillary glucose tesing:
a. Only fingerstick and heelstick for infants are approved sites. The
site will be cleaned, dried, punctured with the Safe T-Pro Lancets.
Wipe the first drop of blood with gauze and use the second free
flowing drop of blood for testing.
b. Do not squeeze the puncture site too hard as it may result in tissue
fluid contamination
c. Alcohol used to cleanse the puncture site for capillary samples
must be thoroughly dry or an error code/inaccurate result may
occur.
3. For Venous, arterial or neonatal blood sample collected in lithium heparin:
a. Use only the blood drawing device assigned to the patient. Do not
share blood drawing devices between patients.
b. Discard blood drawing device appropriately.
B. Patient Testing
1. Using two patient identifiers, verify patient identification, and explain
procedure to patient and/or family.
2. Press the power ON/OFF button to turn on the meter. The meter will
perform Self-Checks.
3. Touch the forward icon [] or wait 5 seconds to automatically proceed to
the Operator ID screen.
4. Press and release the “barcode” button. The button now appears with a
black background during the scan.
5. Hold the meter so that the window of the barcode scanner is
approximately 4 8 inches above the barcode you wish to read. The
meter beeps once the barcode has been read successfully. Operator ID
may also be entered manually by hand-entering your own Employee ID
number.
6. From the Main Menu screen, touch Patient Test.
7. For those sites utilizing patient identification, you must SCAN THE
PATIENT’S BARCODE.
------
------
Rev 1, Ver 2, 9/9/2013