Accu-Chek Go Instrukcja Obsługi Strona 175

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 208
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 174
175
Symbols, error messages and troubleshooting
13
This error message ... means:

O You bent or moved the test strip dur-
ing the test.
Push the ejector down to eject the
test strip. Repeat the test with a new
test strip.
O You inserted a used test strip.
Push the ejector down to eject the
test strip. Repeat the test with a new
test strip.
O You used a test strip that is past its
expiry date.
Push the ejector down to eject the
test strip. Test again with test strips
that have not expired. Remove the old
code chip and insert the one belong-
ing to the new test strips.

Przeglądanie stron 174
1 2 ... 170 171 172 173 174 175 176 177 178 179 180 ... 207 208

Komentarze do niniejszej Instrukcji

Brak uwag